Correlation Between Central Corneal Thickness and Intraocular Pressure with the Goldmann Applanation Tonometer

Authors

DOI:

https://doi.org/10.47672/ajhmn.1681

Keywords:

Central Corneal Thickness, Intraocular Pressure, Relationship, Goldmann Applanation Tonometer

Abstract

Purpose: The aim of this review is to investigate the correlation between central corneal thickness (CCT) and Intraocular pressure with the Goldmann applanation tonometer (GAT).

Materials and Methods: A literature search using Medline database was conducted for the years 1946 to 2022 to explore the relationship between central corneal thickness and Intraocular pressure measured by the Goldmann applanation tonometer. The search was restricted to studies published in English language. The first 80 articles were reviewed and five relevant studies were selected in this analysis.

Findings: Impacts of central corneal thickness on Intraocular pressure were investigated in five studies. Overestimation of intraocular pressure is associated with thick cornea while underestimation of intraocular pressure was associated with thin cornea

Implications to Theory, Practice and Policy: Central corneal thickness can affect the result of Intraocular pressure measurement with the Goldmann applanation tonometer (GAT) . Assessment of central corneal thickness (CCT) is necessary for correct interpretation of IOP measurements obtained with the Goldmann applanation tonometer (GAT).

Downloads

Download data is not yet available.

References

Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol 1975;53:34.

Stamper RL. A history of intraocular pressure and its measurement. Optom Vis Sci. 2011;88(1):E16-E28.

Salvetat ML, Zeppieri M, Tosoni C, Brusini P. Repeatability and accuracy of applanation resonance tonometry in healthy subjects and patients with glaucoma. Acta Ophthalmol. 2014;92(1):e66-e73

Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115:592-6.

Bhan A, Browning AC, Shah S, et al. Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 2002;43:1389-92.

Matsumoto T, Makino H, Uozato H, et al. The influence of corneal thickness and curvature on the difference between intraocular pressure measurements obtained with a non-contact tonometer and those with a Goldmann applanation tonometer. Jpn J Ophthalmol 2000;44:691.

Shimmyo M, Ross AJ, Moy A, et al. Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African Americans. Am J Ophthalmol 2003;136:603-13.

Mardelli PG, Piebenga LW, Whitacre MM, et al. The effect of excimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer. Ophthalmology 1997;104945-8; discussion 949,.

Downloads

Published

2023-12-13

How to Cite

Alshehri, F., & Alrashidi, B. (2023). Correlation Between Central Corneal Thickness and Intraocular Pressure with the Goldmann Applanation Tonometer. American Journal of Health, Medicine and Nursing Practice, 9(5), 23–26. https://doi.org/10.47672/ajhmn.1681