Purpose: The regularly used contrast media in the angiography procedures and interventions are known to carry certain risks to the patient. This includes allergy, nephrotoxicity, and pulmonary edema. Additionally, radiation exposure is associated with high bone marrow depression, infertility, and other hazards. The purpose of the study was to compare the Dynamic Roadmap technology to the regular contrast used in coronary intervention regarding radiation exposure, fluoroscopy time, and incidence of Contrast Induced Nephropathy.

Methodology: Observational prospective cohort with 2 arms where 40 patients were randomly divided into 2 arms a case and control groups. The case group had DRM technology used in their procedures where the control group received the standard contrast used in the National Heart Institute.

Findings: Findings showed no statistically significant difference between the 2 groups in age, sex, and risk profile including status of Hypertension, Diabetes Mellitus, Smoking and history of Ischemic Heart Disease. There was no difference in echocardiographic findings of both groups. There was significant difference in creatinine level at the 5th day of intervention and volume of contrast used and fluoroscopy time. DRM reduced contrast volume, reduced radiation exposure despite not reducing fluoroscopy or procedure time. This resulted in reduction of elevation in serum creatinine levels with similar success rates.

Recommendations: Using the least possible contrast volume and radiation doses should always be target of the operator. Dynamic roadmap technology is recommended in all coronary interventions specially those at high risk of CIN with CKD.

Downloads

Download data is not yet available.

Author Biographies

Mamdouh Maher

Department of Cardiology, the National Heart Institute-Egypt, 5 Ibn El Nafees st, Al Kit- Kat, Giza, 12651.

Bassem Zarif

Department of Cardiology, the National Heart Institute-Egypt, 5 Ibn El Nafees st, Al Kit- Kat, Giza, 12651.

Aya Elgamal

Department of Cardiology, the National Heart Institute-Egypt, 5 Ibn El Nafees st, Al Kit- Kat, Giza, 12651.

Hatem Khairy

Department of Cardiology, the National Heart Institute-Egypt, 5 Ibn El Nafees st, Al Kit- Kat, Giza, 12651.

Ahmed Magdy

Department of Cardiology, the National Heart Institute-Egypt, 5 Ibn El Nafees st, Al Kit- Kat, Giza, 12651.

References

Khan MA, Hashim MJ, Mustafa H, Baniyas MY, Al Suwaidi SKBM, AlKatheeri R, et al. Global Epidemiology of Ischemic Heart Disease: Results from the Global Burden of Disease Study. Cureus. 2020;12(7).

Lun Z, Liu L, Chen G, Ying M, Liu J, Wang B, et al. The global incidence and mortality of contrast-associated acute kidney injury following coronary angiography: a meta-analysis of 1.2 million patients. J Nephrol [Internet]. 2021;34(5):1479–89. Available from: https://doi.org/10.1007/s40620-021-01021-1

Bratschitsch G, Leitner L, Stücklschweiger G, Guss H, Sadoghi P, Puchwein P, et al. Radiation Exposure of Patient and Operating Room Personnel by Fluoroscopy and Navigation during Spinal Surgery. Sci Rep. 2019;9(1):1–5.

Hauptmann M, Daniels RD, Cardis E, Cullings HM, Kendall G, Laurier D, et al. Epidemiological studies of low-dose ionizing radiation and cancer: Summary bias assessment and meta-analysis. J Natl Cancer Inst - Monogr. 2020;2020(56):188–200.

Almendarez M, Gurm HS, Mariani J, Montorfano M, Brilakis ES, Mehran R, et al. Procedural Strategies to Reduce the Incidence of Contrast-Induced Acute Kidney Injury During Percutaneous Coronary Intervention. JACC Cardiovasc Interv. 2019;12(19):1877–88.

Albrecht MH, Vogl TJ, Wichmann JL, Martin SS, Scholtz JE, Fischer S, et al. Dynamic 4D-CT angiography for guiding transarterial chemoembolization: Impact on the reduction of contrast material, operator radiation exposure, catheter consumption, and diagnostic confidence. RoFo Fortschritte auf dem Gebiet der Rontgenstrahlen und der Bildgeb Verfahren. 2018;190(6):513–20.

Piayda K, Phinicarides R, Afzal S, Veulemans V, Jung C, Hellhammer K, et al. TCT-365 Dynamic Coronary Roadmap for Percutaneous Coronary Intervention Effectively Reduces Contrast Medium Exposure: Insights From an Open-Label, Randomized Trial. J Am Coll Cardiol [Internet]. 2021;78(19):B150. Available from: https://doi.org/10.1016/j.jacc.2021.09.1218

Piayda K, Kleinebrecht L, Afzal S, Bullens R, Ter Horst I, Polzin A, et al. Dynamic coronary roadmapping during percutaneous coronary intervention: a feasibility study. Eur J Med Res [Internet]. 2018;23(1):36. Available from: https://doi.org/10.1186/s40001-018-0333-x

Yabe T, Muramatsu T, Tsukahara R, Nakano M, Takimura H, Kawano M, et al. The impact of percutaneous coronary intervention using the novel dynamic coronary roadmap system. Heart Vessels [Internet]. 2020;35(3):323–30. Available from: https://doi.org/10.1007/s00380-019-01502-1

Downloads

Published

2022-12-05

How to Cite

Maher, M. ., Zarif, B. ., Elgamal, A. ., Khairy, H. ., & Magdy, A. . (2022). Dynamic Coronary Roadmap for Contrast, and Radiation Time Reduction during Coronary Intervention (DRM-COR). American Journal of Health, Medicine and Nursing Practice, 7(12), 32 - 39. https://doi.org/10.47672/ajhmn.1298